CHE 450G: Practical Inorganic Chemistry

U of KY Dept of Chemistry

FT-IR Instruction Sheet


This handout explains how to use the Mattson Galaxy 5020 Fourier Transform Infrared spectrometer to collect your data. This is an actual research instrument, not 450G laboratory equipment. Please follow these instructions exactly.


How do I obtain a spectrum?

  1. Prepare the instrument for a background/sample scan:

    1. Sign in on the instrument logbook. Use your last names and "CHE 450G".

    2. Turn up the nitrogen flow (upper black knob located at the middle of the bench. Do not adjust the lower black knob).

        Dry nitrogen gas is continually purging the optics and sample compartment of the instrument. The flow is turned down to a minimum when the instrument is not in use to conserve nitrogen.

    3. If you are using a reference sample (i.e. solution IR etc.), open the sample compartment and place it in the sample holder. Close the lid.

    4. If the instrument is in "Standby" mode, the Mirror Stopped and Standby lights will be lith. Press the Standby button to take it out of this mode.

        There is a mirror inside the instrument that is continuously moving back and forth. By placing the instrument in Standby mode, the mirror is stopped, saving wear on the optics.

    5. After pressing the Standby button, the instrument will do a self test and then the Data Acquisition, Ready and Reference lights will come on (Acquisition will be blinking; this is normal). The display will read "** Ready to collect spectrum**"

  2. Turn on the computer and start the WinFIRST software:

    1. Turn on the computer, printer and monitor (if switched off). DO NOT touch any keys until it has booted and Windows has started.

    2. The computer should go into Windows automatically.

    3. In the Program Manager the WinFIRST directory will usually be open. Double click the WinFIRST icon to start the program.

      Note: If the icon is not visible you will have to open the WinFIRST directory.

  3. Call up the Acquisition Control Panel.

    1. This will normally open automatically. If not you can click on the fourth icon from the left in the toolbar at the top of the screen (it looks like a squiggly line).

    2. The Control Panel will open. It can appear as a large box or a small box depending on whether Options has been clicked. If the box is small, click Options to make it large.

        Note: The default value for scans is 16 scans at 2 cm-1 resolution. The default scan range is 600 - 4000 cm-1. Please do not change these values.

    3. Click on the button Load Method.... You will be presented with a dialog that asks you which method to use. Click on CHE450G method and then click OK.

        This tells the instrument to save your data in the CHE 450G folder in the Data directory. Be sure to pick the proper folder or you may save into someone else's directory! Note: If you are running multiple samples you only have to Load Method the first time.

  4. Perform a Background Scan

    1. Click on the Background radio button on the Control Panel if it is not already selected.

    2. Click on the big green Scan button.

    3. The computer will say "Please Prepare for a Background Scan". Click OK when you are ready to proceed (usually you would like the sample compartment to purge for at least a minute after you close the sample compartment door).

        If nothing happens (i.e. no data collection) then try again from step 2.

    4. Acquisition of the background begins. You will see a status report in the window and in the lower right it will say "xx% complete".

    5. When the computer is done, it will ask for a data file name. Enter something no longer than 8 letters/numbers as you are constrained by DOS. Do not use any file extensions (periods) in your name. Click OK.

        NOTE: It is highly recommended that you name your files by your initials and notebook page. For example: Rob Toreki would have a sample RT22A.

        You can use the same data file name for background and sample scans (they have different file extensions so it's OK).

    6. You will then be prompted for a title. Since this is a background scan, it does not really matter what you type. Click OK.

      The background file is always saved. This is good if you want to run several samples but not rerun the background each time. In usual practice you can throw away the background files (*.bkg) when you are done.

    7. Your Background scan will appear. It typically contains a broad curve and may or may not have large peaks due to CO2 and water depending on how long you purged the sample compartment.

  5. Collect Sample Data

    1. Open the sample compartment and put in your sample. Let the compartment purge for roughly the same amount of time you did for the background.

    2. Click on the Control Panel if it is visible. If not you can click on the multicolored icon in the lower left that says "Control Panel" or on the fourth icon from the left on the toolbar.

    3. Select Sample by clicking on the radio button on the Control Panel.

    4. Click the big green Scan button.

    5. As before, messages will appear and the % complete will be displayed.

    6. You will be prompted with a filename. Enter something no longer than 8 letters/numbers as you are constrained by DOS. Do not use any file extensions (periods) in your name. Click OK.

    7. You will be prompted for a title. It is best to erase everything that is there and put a short description of the sample, date, time and notebook page. Include whether it was a Nujol mull, pellet, solution (and solvent) etc. since you may not remember this information months from now.

    8. Your spectrum will now be displayed. You can enlarge it to full screen size by clicking on the up arrow in the upper right hand corner of the sample window.


How do I obtain a solution spectrum?

  1. Obtain a spectrum of your solvent using the procedure outlined above, save it to disk and then close the file.

  2. Obtain a spectrum of your solvent + sample as above. Be sure to use the same cell (or a matched cell). You can reuse the same background, so no need to recollect that.

  3. Under the File menu select Load Reference.... Select the name of your solvent file and click OK.

  4. We now want to subtract our reference (pure solvent) from our sample (sample + solvent). To do this, go to the Math menu and choose Subtract....

  5. You will see three spectra. The bottom one (in white) is the difference spectrum. Move the slider bar or type a number in the box (something close to 1.0 usually works best) until the difference spectrum is flat as possible in the regions where your solvent peaks normally occur. Be careful not to "oversubtract" when, for example, you have CH peaks from your compound superimposed on the CH peaks of your solvent.

  6. When everything is to your satisfaction, click Done. You may want to change the text header on your file to include "solvent subtracted" and then save the file under a new name.

How do I manipulate the spectrum and massage the data?

This is not meant to be a complete guide.